Samsung SONOACE R3 Manuel d'utilisateur Page 144

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 308
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 143
User Manual
5-20
2D Hip Joint
This is a basic measurement that is available in all diagnosis modes. You can measure the angle of two
straight lines to a straight line in a 2D image. This measurement is commonly used in the obstetrics
application.
1. Press the Softmenu button [4] Angle to select 2D Hip Joint. “2D Hip Joint is displayed on the left
side of the screen.
2. Draw three straight lines. The method for measuring a line is the same as in “2D Distance (Line).
3. The angle between them will be calculated automatically.
X
α: The angle between the rst and second straight lines.
X
β: The angle between the rst and third straight lines.
4. When the measurement is nished, its result is shown on the screen. The angles α and β, and the
Hip Joint Type are displayed on the screen.
Please refer to the table below for Hip Joint Type information:
Type α β
1a
60 ≤ α < 90
0
<
β
<
55
1b
60 ≤ α < 90
55 ≤ β
<
90
2a/b
50 ≤ α < 60
0
<
β
<
90
2c
43 ≤ α < 50
77 ≤ β
<
90
d
43 ≤ α < 50
0
<
β
<
77
3/4
0 < α < 43
[Table 5.2 Hip Joint Type Table]
Vue de la page 143
1 2 ... 139 140 141 142 143 144 145 146 147 148 149 ... 307 308

Commentaires sur ces manuels

Pas de commentaire