SAMSUNG MEDISON DIAGNOSTIC ULTRASOUND SYSTEMSONOACE R3User ManualVolume 1
XMedical Electrical Equipment, Part 1-4: General Requirements for Safety - Collateral Standard: Programmable Electrical Medical Systems [IEC 60601-1-
User Manual4-12 2D Mode SoftmenuFollowing items are displayed on soft menu in 2D mode. Select the item and adjust the value with the Softmenu button
Chapter 4 Diagnosis Modes4-13M ModeBy drawing the M line across a desired area within the 2D image, you can view changes over time.This is useful for
User Manual4-14 M Mode MenuFollowing items are displayed on menu when you press the Menu dial-button in M mode. Select the item and adjust the value
Chapter 4 Diagnosis Modes4-15Color Doppler ModeThis mode displays the colored blood ow pattern of the ROI (Region of Interest) within the 2D image.
User Manual4-16Use the Change button to adjust the position and size of the ROI box. Each time you press the Change button, the status of the ROI box
Chapter 4 Diagnosis Modes4-17 Color modeSet the contents that are displayed in color. Select Vel., Power, Var., or VelVar. with the Menu dial-button
User Manual4-18▶Tag Width: It is available when Color Tag is on. Use the Menu dial-button to select the range of tags, from between 8 - 256. Tag Widt
Chapter 4 Diagnosis Modes4-19Power Doppler ModeThis mode displays the color intensity of blood ow within the ROI in the 2D image. It is appropriate
User Manual4-20 ROI BoxROI stands for Region Of Interest. In Power Doppler Mode, ROI indicates the area of the 2D image where color (blood ow) infor
Chapter 4 Diagnosis Modes4-21PW Spectral Doppler ModePW stands for Pulse Wave. PW Spectral Doppler Mode gives information on the speed of blood flow
Precautions for UseYou should be familiar with all of these areas before attempting to use this manual or your ultrasound system. Please keep this
User Manual4-22 Sample VolumeWhen Sample Volume is positioned over the blood flow in a 2D image, the Doppler Spectrum is displayed. Information on th
Chapter 4 Diagnosis Modes4-23[Figure 4.8 HPRF]NOTE: HPRF cannot be activated in Simultaneous Mode. Also, HPRF cannot be activated if the PRF value X
User Manual4-24 Spectrum Ehc.Spectrum Ehc stands for Spectrum Enhancement. This functionis used to compensate the signal of low gain Spectral Doppler
Chapter 4 Diagnosis Modes4-25 PW Spectral Doppler Mode Soft MenuFollowing items are displayed on soft menu in PW Spectral Doppler mode. Select the
User Manual4-26 Combined Modes2D/C/PW ModeColor Doppler Mode and PW Spectral Doppler Mode are displayed simultaneously. Press the PW button on the co
Chapter 4 Diagnosis Modes4-27 Changing Combined Mode Format Changing Active ImagePress the Set button. The currently active image mode, such as ‘P
User Manual4-28 Dual ModeThe multi-image viewing function is only available when the images and conditions on both sides are identical. The current a
Chapter 4 Diagnosis Modes4-29 3D Mode3DThe 3D Mode displays tissues or objects in the human body as three-dimensional images inste-ad of two-dimensi
User Manual4-30Acquiring a 3D Image The following explains how to acquire a 3D image. 1. Press the 3D key on the keyboard.2. The ROI box appears. Set
Chapter 4 Diagnosis Modes4-313D View3D images can be seen through the 3D View application.[Figure 4.9 3D View] Viewing a 3D Image ▶ Select the menu
DANGER: Describes precautions necessary to prevent user hazards of great urgency. Ignoring a DANGER warning will risk life-threatening injury.WARNING:
User Manual4-32 3D Image Optimization Threshold (Low, High) AdjustmentAdjust the slide bar of the threshold using the Trackball and the Set button.
Chapter 4 Diagnosis Modes4-33[Figure 4.10 Soft Menu - Rotation Mode]Tips!Other ways to rotate the image1. The ROT slide bar at the bottom of the scr
User Manual4-34 MagnicationTo magnify or reduce the image size, enter Magnication mode by pressing the Magnication button, and move the Trackball
Chapter 4 Diagnosis Modes4-35 Multi planar Press Multi planar, and 3D multiplanar screen appears. Shows volume data in cross-section images of Coron
User Manual4-36Use X/Y/Z slide on the screen to rotate reference plane.AYYZXXYZZXCB[Figure 4.17 The standard coordinates systems of Multi-planar]YXZRe
Chapter 5Measurements and CalculationsChapter5 Measurements and CalculationsMeasurement Accuracy ...5-3Causes of Me
Chapter 5Report ... 5-51Viewing Report ...
Chapter 5 Measurements and Calculations5-3 Measurement AccuracyMeasurement values can vary, depending on the nature of the ultrasound, the body’s re
User Manual5-4 Ultrasound VelocityXThe velocity of ultrasound used during measurement is usually 1,540 m/s on average.XThe velocity of ultrasound
Chapter 5 Measurements and Calculations5-5Optimization of Measurement Accuracy 2D ModeXResolution is in proportion to the frequency of the probe.
Revision HistoryThe revision history of this manual is as follows.VERSION DATE NOTEv2.01.03-00 2015.04.14Initial ReleaseSystem Upgrades and Manual Se
User Manual5-6 Color/Power Doppler ModeXA protocol is not specied for images in Color Doppler Mode or Power Doppler Mode. Therefore, the same limi
Chapter 5 Measurements and Calculations5-7Measurement Accuracy TableThe following tables show the accuracy of the measurements available using the pr
User Manual5-8 PW Spectral Doppler ModeDoppler MeasurementSystem Tolerance(whichever is greater)Methodology Range(1)Velocity < ± 15% Phantom 0.
Chapter 5 Measurements and Calculations5-9 Basic MeasurementsPress the Caliper button on the control panel. NOTE: Take basic measurements of distanc
User Manual5-10 Basic Measurement OperationsThe following is the information on common button operations for basic measurements:[Figure 5.1 Basic Me
Chapter 5 Measurements and Calculations5-11Distance Measurement 2D Distance (Line) This is a basic measurement that is available in all diagnosis m
User Manual5-12 M DistanceThis is a basic measurement that is available in M Mode only. You can specify two points in an M image and measure the dis
Chapter 5 Measurements and Calculations5-13The equations used for D Velocity measurement are as follows: Doppler Velocity (A/B)This is a basic meas
User Manual5-14 Doppler TraceThis is a basic measurement that is available in Spectral Doppler Mode only. You can specify a point in a Spectral Dopp
Chapter 5 Measurements and Calculations5-15Circumference and Area Measurement 2D EllipseThis is a basic measurement that is available in all diagn
User Manual5-16 2D TraceThis is a basic measurement that is available in all diagnosis modes. You can measure the circumference and area of an irreg
Chapter 5 Measurements and Calculations5-17Volume Measurement 3 DistanceThis is a basic measurement that is available in all diagnosis modes. You c
User Manual5-18 Distance + EllipseThis is a basic measurement that is available in all diagnosis modes. You can measure the volume of an ob-ject in
Chapter 5 Measurements and Calculations5-19Angle Measurement 2D Angle This is a basic measurement that is available in all diagnosis modes. Specify
User Manual5-20 2D Hip Joint This is a basic measurement that is available in all diagnosis modes. You can measure the angle of two straight lines t
Chapter 5 Measurements and Calculations5-21 Calculations by Application Press the Calc button on the control panel.Things to note Before Taking Me
User Manual5-22 Measurement OperationsThe following gives information on the common button operations for measurements: Select/Change ApplicationUs
Chapter 5 Measurements and Calculations5-23 Soft Menu[Figure 5.2 Calculations by Application – Soft Menu] Menu A/Menu BWhen there are a lot of mea
User Manual5-24Common Measurement MethodsThis section provides information on the common measurement methods used for applications. Measurements in
Chapter 5 Measurements and Calculations5-25Tips!Things to consider for Doppler Spectrum Auto TraceThe state of a Doppler spectrum may affect measureme
Table of Contents 15Table of Contents – Volume 1Chapter1 SafetyPurpose of Use ...
User Manual5-26 Volume Flow MeasurementSelect Volume Flow in the Measurement menu. Volume Flow allows you to measure and calculate an area or distance
Chapter 5 Measurements and Calculations5-27 Heart Rate Measurement HR (Heart Rate)You can calculate heart rates for a certain period of time.1. Se
User Manual5-28OB Calculations Before Obtaining OB CalculationsSet GA Equation, GA Table and OB Measurement menus. The user can manually write, back
Chapter 5 Measurements and Calculations5-29Measurement Menu Item Mode Method UnitFetal BiometryGS All Distance cm, mmCRL All Distance cm, mmYS All Di
User Manual5-30Measurement Menu Item Mode Method UnitFetal OthersFoot All Distance cm, mmEar All Distance cm, mmMP All Distance cm, mmAFIAll All Dista
Chapter 5 Measurements and Calculations5-31 Automatic Calculation Some items in the measurement menu are automatically calculated based on measurem
User Manual5-32 AFI (Amniotic Fluid Index)This measures the amniotic uid index. Measurements are performed by dividing the pregnant woman’s abdomen
Chapter 5 Measurements and Calculations5-33GYN CalculationsThe measurement method for each item is the same as for basic measurement. Measured items
User Manual5-34Measurement Menu Item Mode Method UnitRt./Lt. Follicle 1 ~ 12 All Distance cm, mmRt./Lt. Ovarian AAuto Trace PW Doppler Spectrum TraceL
Chapter 5 Measurements and Calculations5-35Cardiac CalculationsNOTE: XCardiac measurement is an optional item for SONOACE R3. XSince the cardiac bl
User Manual16Chapter3 Starting Diagnosis Power Supply ...
User Manual5-36Measurement Menu Item Mode Method Unit2D MeasureAll All Continuous cm, mmIVSd All Distance cm, mmIVSs All Distance cm, mmLVDd All Dist
Chapter 5 Measurements and Calculations5-37Measurement Menu Item Mode Method UnitMV (M)All Points M ContinuousD-E M Distance cm, mmE-F Slope M Velo
User Manual5-38Measurement Menu Item Mode Method UnitLVOT DopplerAuto Trace PWDoppler Spectrum TraceLimited Trace PWDoppler Spectrum TraceManual Trace
Chapter 5 Measurements and Calculations5-39Measurement Menu Item Mode Method UnitTV InowLimited Trace PWDoppler Spectrum TraceManual Trace PWDoppler
User Manual5-40Measurement Menu Item Mode Method UnitPV RegurgAuto Trace PWDoppler Spectrum TraceLimited Trace PWDoppler Spectrum TraceManual Trace PW
Chapter 5 Measurements and Calculations5-41Measurement Menu Item Mode Method UnitQp:QsSys. VTI PW Distance cm, mmSys. HR PW Heart Rate bpmSys. Dia
User Manual5-42Vascular Calculations It is convenient to calculate each measurement value on the Spectral Doppler image.The measurement method for eac
Chapter 5 Measurements and Calculations5-43Measurement Menu Item Mode Method UnitRt./ Lt. SubclavianAuto Trace PW Doppler Spectrum TraceLimited Trace
User Manual5-44Urology CalculationsThe measurement methods of each menu vary with the Volume Method set at Setup > Measure > Urology. The measur
Chapter 5 Measurements and Calculations5-45Measurement Menu Item Mode Method UnitProstate Vol.All AllCalculated after distance measurementml1st Dia.
Table of Contents 17Chapter5 Measurements and CalculationsMeasurement Accuracy ...
User Manual5-46 EllipsoidVolume is calculated using the Main diameter and Beside Diameter values.Measurement Menu Item Mode Method UnitVolume FlowAut
Chapter 5 Measurements and Calculations5-47 Sum of 20 DisksMeasure the prostate circumference and then use the trackball and the Set button to measu
User Manual5-48Fetal Echo CalculationsThe measurement method for each item is the same as for basic measurement. In addition, measurement items are si
Chapter 5 Measurements and Calculations5-49Measurement Menu Item Mode Method Unit2D EchoAsc. Aorta Distance Doppler Spectrum Trace cm, mmMPA Distan
User Manual5-50Measurement Menu Item Mode Method UnitIVC Same as above Same as aboveDuct. Ven Same as above Same as aboveAsc. Aorta Same as above Same
Chapter 5 Measurements and Calculations5-51 ReportCheck the measurement and calculation results in the report of each application. ▶ Press the Report
User Manual5-52Viewing Report▶ The measurement and calculation items and their values are displayed. ▶ When the report screen has more than two pages,
Chapter 5 Measurements and Calculations5-53 Open Line TransferTransfer the report to another computer or printer. Press the Open Line Transfer on t
User Manual5-54Graph Function▶ This function is used to display the GA values such as BPD, HC, AC and FL, allowing the user to diagnose fetal growth.
SAMSUNG MEDISON DIAGNOSTIC ULTRASOUND SYSTEMSONOACE R3User ManualVolume 2
SAMSUNG MEDISON DIAGNOSTIC ULTRASOUND SYSTEMVersion 2.01EnglishSONOACE R3User Manual
Table of Contents 5Table of Contents – Volume 2Chapter6 Image ManagementCINE/LOOP ...
User Manual6Setting DICOM (Optional)...7-
Table of Contents 7Chapter9 ProbesProbes ...
Chapter 6Image ManagementChapter6 Image ManagementCINE/LOOP ... 6-3Annotating Images ...
Chapter 6 Image Managemen6-3 CINE/LOOPImages are automatically saved in the memory during scanning. The saved images can be useful when a pa-tient is
Chapter1Purpose of Use ... 1-3Contraindications ...
User Manual6-4 How to Review ImagesUse the trackball on the control panel. You can search for saved images in chronological order by rotating the Trac
Chapter 6 Image Managemen6-5 Cine SaveUse the Softmenu button [4]. Save the Cine whole Cine images. NOTE: XWhen saving a Cine image, the current ti
User Manual6-6 Annotating Images TextYou can enter comments or text over an image. This function can be useful when the diagnosis area is dier-entia
Chapter 6 Image Managemen6-7 AutotextThis function allows you to enter text automatically by using an abbreviation. In this way, you can enter text e
User Manual6-8BodyMarker You can place a BodyMarker over an image. This function can be useful when the diagnosis area is dierenti-ated or displayed.
Chapter 6 Image Managemen6-9 Entering BodyMarker 1. Press the BodyMarker key on the keyboards, BodyMarkers will appear on the user information area.
User Manual6-10Indicator You can place an Indicator over an image. This function can be useful when the diagnosis area is dierentiat-ed or displayed.
Chapter 6 Image Managemen6-11 Deleting Indicator Press the Clear button on the control panel and all Indicators displayed on the screen will be delet
User Manual6-12 Saving and Transferring ImagesSaving Images WARNING: You must enter the patient ID because all images are saved according to patient
Chapter 6 Image Managemen6-13Transferring Images Send images using DICOM. For more information on the DICOM server settings and DICOM operation, see
User Manual6-14 Printing and Recording ImagesPrinting ImagesPress the Print button on the control panel to print out a scanned image using a printer.
Chapter 6 Image Managemen6-15 SONOVIEWSONOVIEW is an integrated image management program in SONOACE R3. It allows you to save, view and de-lete image
User Manual6-16Exam ViewNOTE: If you select an icon on the Exam View screen, the name of icon will be displayed. Starting Exam ViewPress Exam Search
Chapter 6 Image Managemen6-17 Selecting Exam Use the trackball and the Set button to select an examination. To cancel the exam selection, press the S
User Manual6-18 Exam Export Multiple exams can be selected and exported simultaneously.1. Select exam(s) and press Export to bring up the Export windo
Chapter 6 Image Managemen6-19Tips!When selecting 2D Cine, an image and a video are saved together. The saved AVI file (*.avi) can be played by a video
User Manual6-20 Using USB Flash MemoryUSB ash memory can be connected to the USB port on the left side or the rear panel. Only one ash memory can b
Chapter 6 Image Managemen6-21[Figure 6.9 DICOM Send] Printing Exam This function uses printer(s) on the DICOM network. DICOM Print will not work if t
User Manual6-22Exam ReviewSaved images can be managed using measurement, printing, sending and many other functions. [Figure 6.10 Exam Review]NOTE: If
Chapter 6 Image Managemen6-23 Close ExamCloses the current exam(s). Exam SearchThe Exam View screen appears. You can browse through exams to select,
Chapter 1 Safety 1-3 Purpose of UseThe SONOACE R3 Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and
User Manual6-24 PrintingThis function prints images or simple reports.1. Press Print from the Exam Review screen. 2. Select settings from the Print wi
Chapter 6 Image Managemen6-25 DICOM Printing (Optional)This function uses DICOM to print selected images. This button is deactivated for products tha
User Manual6-26 DeleteSelect image(s) and press Delete Image on the Exam Review screen. To delete many images at once, use Trackball and Set button du
Chapter 6 Image Managemen6-274. Use the trackball to move the cursor to another location and press the Set button to set the end point. ▶ Press the C
Chapter 7UtilitiesChapter7 UtilitiesSetting System ... 7-3General ...
Chapter 7Setting Option ...7-34Setting Peripheral Devices ...7-35Information ..
Chapter 7 Utilities7-3 Setting System This mode is used for system settings. It does not aect image output. The setup may be modied depending on s
User Manual7-4 InstitutionEnter the name of the hospital/institution. LanguageThis sets the language to be used. English, Deutsch, Français, Italian
Chapter 7 Utilities7-5Tips!How to set the date and time.[Figure 7.2 Date & Time Setup]XClick Set in the Title tab of System menu.XSet the Date/
User Manual1-4 Safety SignsPlease read this chapter before using the Samsung Medison ultrasound system. It is relevant to the ultrasound system, the
User Manual7-6Tips!How to edit the Autotext Set.Click Edit on the screen. The Autotext Table Edit Table window will be appeared. To add a new abbrevia
Chapter 7 Utilities7-7Display To set the information about images and related data, select the Display tab in the System menu. Auto FreezeAfter the
User Manual7-8 Display GenderThis sets whether or not the gender of the patient is displayed. If ‘On’ is selected, the patient’s gender is shown. Si
Chapter 7 Utilities7-9Misc.Select the Misc. tab in the System menu.[Figure 7.5 System Setup – Misc.] OB Measure Shortcut KeySelect On to start a des
User Manual7-10 ThumbnailTo use thumbnail list, set this item as ‘On’. Thumbnails of saved images are showed up on the right side of the screen. Se
Chapter 7 Utilities7-11 Setting Measurements Select Measure in the Setup menu to set up measurement functions.1. Press the Setup key on the keyboard
User Manual7-12 GeneralSelect General tab. Measure Item Display CountSelect the number of measurement items to be between 1 and 4. If you select ‘Hid
Chapter 7 Utilities7-13 Doppler Select the Doppler tab in the General tab. [Figure 7.7 Measure - Doppler] Y-Axis UnitThis sets the vertical (Y) axis
User Manual7-14 Mean Trace This menu species the use of Mean Trace Line on the screen for Auto trace/Limited trace.Tips!How to change Doppler settin
Chapter 7 Utilities7-15 ReportSelect the Report tab in the General tab.Select items to mark the header on the report. The header will be shown only t
Chapter 1 Safety 1-5Symbols Description Symbols DescriptionPower on Microphone PortPower o Probe portPower ON for part of the productProtected agai
User Manual7-16OBSelect the OB tab in the Measure menu.[Figure 7.10 Measure – OB] Display SD ValueThis menu sets whether SD (Standard Deviation) is d
Chapter 7 Utilities7-17▶Auxiliary Unit: This sets the auxiliary unit of fetal weight measurement. Unit sets the available auxiliary unit. For instan
User Manual7-18 GraphThe user can select graphs to mark on the OB report or print. Press Graph button and select items. After selecting/deselecting a
Chapter 7 Utilities7-19 Backup/Restore▶BackupIf you want to back up a new user preset to Mo Media, Flash Memory or a HDD, press Backup/Restore on t
User Manual7-20Fetal Echo Select the Fetal Echo tab in the Measure menu.[Figure 7.14 Measurement Setting– Fetal Echo]During obstetric measurement, Fet
Chapter 7 Utilities7-21Cardiac Select the Cardiac tab in the Measure menu. 2D Circumference and Area MethodSelect either the ‘Ellipse’ or ‘Trace’ c
User Manual7-22Urology Measurement SetupSelect the Urology tab in the Measure menu. [Figure 7.16 Measure– Urology] Volume Method▶3 Distance: The vol
Chapter 7 Utilities7-23NOTE: 3 Distances: A = 1st Dia. ; B = 2nd Dia. ; C = 3rd Dia. The default factor value (F) is ‘0.5’. If it is changed, a value
User Manual7-24Vascular Select the Vascular tab in the Measure menu. You can set the peak velocity used in the calculation of the ‘A/B Ratio’ and ‘ICA
Chapter 7 Utilities7-25 Setting DICOM (Optional)Select the DICOM tab in the Setup menu. This function is used to set up the DICOM server and other D
User Manual1-6Labels Phrases containing the words ‘warning’ and/or ‘caution’ are displayed on the product’s surface in order to protect it.
User Manual7-26Setting DICOM InformationEnter the DICOM information for the product in use. For automatic DICOM transmission, select ‘Acquisition in P
Chapter 7 Utilities7-27 ServiceSelect the kind of server to use. The supported DICOM servers are Store, Print, and Worklist. AliasEnter the name of
User Manual7-28[Figure 7.19 DICOM Configuration-Store] Print Server InformationSelect PRINT under Services. Congure the Print Service using DICOM. NO
Chapter 7 Utilities7-29 Border DensitySpecify the border density of an image to print. Select Black or White. Empty DensitySpecify the background c
User Manual7-30 Conguration InfoSpecify the unique value for a printer. Please refer to the DICOM Conformance Statement for the printer.[Figure 7.20
Chapter 7 Utilities7-31 Scheduled Station AE TitleSets the range of AE Title to fetch from the hospital’s Worklist server.▶All: Obtains the list of
User Manual7-32Editing the DICOM Server Information Click Edit in the setting DICOM window to edit the server information already entered.Deleting DIC
Chapter 7 Utilities7-33 DICOM LogView, copy, or delete the DICOM log les.▶View Selected File: Select the log le from ‘Select log les to copy’ an
User Manual7-34 Setting OptionThe S/W serial No information of the system is shown in this window. You can select/cancel S/W options. A user cannot m
Chapter 7 Utilities7-35 Setting Peripheral DevicesThe following describes how to set up the video output type, video input type, printer, foot switc
Chapter 1 Safety 1-7 Electrical SafetyThis equipment is categorized as a Class I device with Type BF applied parts.CAUTION: As for US requirement,
User Manual7-36 Printer 1Select a printer to use. After connecting a USB printer to the USB port of the system, select the printer type on the screen
Chapter 7 Utilities7-37 InformationThe information menu displays information about the system S/W version. Select the Information tab in the Setup me
User Manual7-38 Utility Press Utility key on the keyboard. The Utility menu appears at the left side of the screen. You can select a menu with the Men
Chapter 7 Utilities7-39 Editing Biopsy guide lineNOTE: XNote that the Biopsy Guide Line cannot be edited when the Trapezoidal function is in use for
User Manual7-40 Starting and Finishing BiopsyNOTE: The biopsy guideline must be adjusted before a biopsy can be performed.1. Select On in the Biopsy m
Chapter 7 Utilities7-41Monitor CalibrationSelect Monitor Calibration in Utility menu. Set the color or brightness of the image. [Figure 7.28 Monitor
User Manual7-42PresetPreset allows the user to adjust Image Settings and select a Setting to apply to the system.Select Preset in the utility menu.[Fi
Chapter 7 Utilities7-43 HistogramShows the brightness level of the selected area in a histogram.1. Select Histogram in the Miscellaneous menu.2. Spec
Chapter 8MaintenanceChapter8 MaintenanceOperating Environment ...8-3System Maintenance ...
User Manual1-8CAUTION: XThe system has been designed for 100-120VAC and 200-240VAC; you should select the input voltage of printer and VCR. Prior to
Chapter 8 Maintenance8-3 Operating EnvironmentWhen installing the console, please consider the following:CAUTION: Placing the system near generators
User Manual8-4 System MaintenanceCleaning and disinfections Using inappropriate detergent or disinfectant may damage the product. Please read the fol
Chapter 8 Maintenance8-5 DisinfectionsCAUTION: Use only recommended disinfectants on system surfaces.A disinfectant qualified by the FDA 510 (k) proc
User Manual8-6 Data MaintenanceCAUTION: You may lose information files on user settings or patients, because of shock on the product or internal error
Chapter 8 Maintenance8-7SoftwareChanges can be made to the product software to make it suitable for a particular user’s environment. Howev-er, these
Chapter 9ProbesChapter9 ProbesProbes ...9-3Ultrasound transmission Gel ...
Chapter 9 Probes9-3 ProbesThe probe is a device that sends and receives ultrasound for acquiring image data. It is also called a Transducer or Scanhe
Chapter 1 Safety 1-9ESDElectrostatic discharge (ESD), commonly referred to as a static shock, is a naturally occurring phenomenon. ESD is most preva
User Manual9-4NOTE: XBesides, user can have or select own preset between User1-5. XRefer to ‘Chapter3. Starting Diagnosis’ for how to set probe and
Chapter 9 Probes9-5Probe Application HAR PI SCI DMR+ Q Scan ECGL5-12/60Breast x x x x O xSmall Parts x x x x O xVascular x x x x O xMusculoskeletal x
User Manual9-6Probe Application CM TDI PD TDW CWC2-4/20Abdomen x x O x xCardiac O x O x xC2-5Abdomen x x O x xOB x x O x xGynecology x x O x xFetal He
Chapter 9 Probes9-7Probe Application CM TDI PD TDW CWLE5-12Breast x x O x xSmall Parts x x O x xVascular x x O x xMusculoskeletal x x O x xLN5-12Brea
User Manual9-8 TI TableTI (Thermal Index) is displayed on the title of the screen and shows the estimate of the temperature increase in specic body
Chapter 9 Probes9-9Ultrasound transmission GelUsing an inappropriate ultrasound gel may damage the probe. For proper transmission of the acoustic bea
User Manual9-10 Installing the Sheath1. Use sterile gloves.2. Unpack the sheath and ll it with acoustic coupling gel.3. Insert the probe into the
Chapter 9 Probes9-11 Use and Infection Control of the ProbeWARNING: No neurosurgical treatments or examinations should be carried out on a patient wi
User Manual9-12Cleaning and Disinfecting the ProbeWARNING: XAlways use protective eyewear and gloves when cleaning and disinfecting probes.XInspect
Chapter 9 Probes9-13NamesDisinfectantsCidex OPA 2), 3), 6), 7)Cidex Plus 2), 7)Metricide 2), 7)Omnicide (28)Omnicide 14NSOmnicide - FG2NucleanWavicid
User Manual1-10EMC The testing for EMC(ElectroMagnetic Compatibility) of this system has been performed according to the international standard for EM
User Manual9-14NamesDisinfectants CleanerVirkonSporoxSporox IIGigaseptGigasept AF 3)Gigasept FFHibitanePeraSaveEnzolAlkazymeCidezymeKlenzymeIsopropyl
Chapter 9 Probes9-15NamesCleaner GelEthanol 75%MetrizymeMcKessonNatural ImageAquasonics 100 3)GE Ultrasound Contact GelClear ImageKendallScanWaveleng
User Manual9-16Tips!Symbols (1) Compatible but no EPA Registration (2) FDA 510 (k) cleared (3) Has CE mark (4) Discontinued (5) Under D
Chapter 9 Probes9-17Following is information about manufacturer (or Distributor) of Detergent, Disinfectant, and Ultrasound Gel. Product Manufacturer
User Manual9-18 CleaningCleaning is an important procedure that is carried out before disinfecting the probe. The probe must be cleaned after each use
Chapter 9 Probes9-19 DisinfectionOnly disinfect vaginal and rectal probes. A 10-6 reduction in pathogens should be reached following the disinfection
User Manual9-20 BiopsyA biopsy is an examination method that surgically extracts tissue from the patient for examination. The probe and the biopsy kit
Chapter 9 Probes9-21Using the Biopsy KitWARNING: XVerify the condition of the biopsy needle before use. Do not use a bent biopsy needle.XThe biopsy
User Manual9-225. Insert the needle into the needle guide. Perform the puncture by sliding the needle through the groove in the guide until the needl
Chapter 9 Probes9-23Cleaning and Disinfecting Biopsy KitWash and disinfect the biopsy kit to reduce pathogens to the level of 10-6. Some components o
Chapter 1 Safety 1-11 Approved Cables, Probes and Peripherals for EMC CablesCables connected to this product may aect its emissions. Refer to th
User Manual9-24 Cleaning and Disinfection of plastic biopsy kit Cleaning 1. After use, remove the biopsy kit from the probe.2. Disassemble the bio
Chapter 9 Probes9-25Assembling the Biopsy Kit C2-5 Biopsy Guide (Plastic) Biopsy SpecicationsModel NameBP-KIT-045ComponentBiopsy BracketNeedle Gu
User Manual9-263. Mount the needle guide for the desired gauge onto the adaptor.4. Install the needle into the needle guide.
Chapter 9 Probes9-27 C2-5 Biopsy Guide (Metal) Biopsy SpecicationsModel nameBP-KIT-042 ComponentBiopsy BracketNeedle Guide (16G)Needle Guide (18G
User Manual9-283. Mount the needle guide for the desired gauge onto the adaptor.4. Install the needle into the needle guide.
Chapter 9 Probes9-29 C2-8 Biopsy Guide Biopsy SpecicationsModel nameBP-KIT-035ComponentBiopsy BracketNeedle Guide1 EA24 EAMaterialAcetal Copolyme
User Manual9-303. Mount the needle guide for the desired gauge onto the adaptor.4. Install the needle into the needle guide.
Chapter 9 Probes9-31 CN2-8 Biopsy Guide Biopsy SpecicationsModel nameBP-KIT-038ComponentAdapterNeedle Guide (16G)Needle Guide (18G)Needle Guide (
User Manual9-323. Mount the needle guide for the desired gauge onto the adaptor. 4. Install the needle into the needle guide.
Chapter 9 Probes9-33 L5-12/60 Biopsy Guide Biopsy SpecicationsModel nameBP-KIT-039ComponentAdapterNeedle Guide (16G)Needle Guide (18G)Needle Guid
User Manual1-12Immunity test IEC 60601 Test level Compliance levelElectromagnetic environment -guidanceElectrotatic discharge (ESD)IEC 61000-4-2±6KV C
User Manual9-343. Secure the Needle Guide Clip. 4. Insert the needle into the needle guide for use.
Chapter 9 Probes9-35 LE5-12 Biopsy Guide Biopsy SpecicationsModel nameBP-KIT-040ComponentBiopsy Adapter 1 EAMaterialAcetal CopolymerAvailable Ga
User Manual9-363. Secure the Needle Guide Clip. 4. Insert the needle into the needle guide for use.
Chapter 9 Probes9-37 LN5-12 Biopsy Guide Biopsy SpecicationsModel nameBP-KIT-040ComponentBiopsy Adapter 1 EAMaterialAcetal CopolymerAvailable Ga
User Manual9-383. Secure the Needle Guide Clip.4. Insert the needle into the needle guide for use.
Chapter 9 Probes9-39 LN5-12/40 Biopsy Guide Biopsy SpecicationsModel nameBP-KIT-036ComponentAdapterNeedle Guide1 EA24 EAMaterialAcetal CopolymerA
User Manual9-403. Secure the Needle Guide Clip.4. Insert the needle into the needle guide for use.
Chapter 9 Probes9-41 EC4-9 Biopsy guide Biopsy SpecicationsModel nameBP-KIT-024ComponentBiopsy AdapterSheath1 EA2 EAMaterialStainlessAvailable Gau
User Manual9-42 EVN4-9 Biopsy Guide Biopsy SpecicationsModel nameBP-KIT-024ComponentBiopsy AdapterSheath1 EA2 EAMaterialStainlessAvailable Gauges1
Chapter 1 Safety 1-13Immunity testIEC 60601Test levelCompliance levelElectromagneticenvironment - guidanceConducted RFIEC 61000-4-63 Vrms150 kHz to
User Manual1-14 Recommended separation distances between portable and mobile RF communications equipment and the SONOACE R3This product is intended
Chapter 1 Safety 1-15 Avoiding Electromagnetic InterferenceTypical interference on Ultrasound Imaging Systems varies depending on Electromagnetic
User Manual1-16X Is the interference intermittent or constant?X Does the interference show up only with one transducers operating at the same freque
Chapter 1 Safety 1-17 Mechanical SafetySafety NoteCAUTION: XDo not apply excessive force to the product. XBe sure to install the console on a sta
User Manual1-18 Biological SafetyVerify the alignment of the biopsy guide before use. See the “Chapter 9. Probes” section of this manual.WARNING: XU
Chapter 1 Safety 1-19 Applying ALARAThe system-imaging mode used depends upon the information needed. 2D-mode and M-mode imaging provide anatomica
User Manual1-20 Indirect ControlsThe indirect controls are those that have an indirect effect on acoustic intensity. These controls affect imaging m
Chapter 1 Safety 1-21 Additional ConsiderationsEnsure that scanning time is kept to a minimum, and ensure that only medically required scanning is
User Manual1-22 Mechanical Index (MI) DisplayMechanical bioeects are threshold phenomena that occur when a certain level of output is exceeded. The
Chapter 1 Safety 1-23The measured water tank values are de-rated using a conservative, industry standard, attenuation coecient of 0.3dB/cm-MHz.Cons
User Manual1-24 2D-mode Controls 2D-mode sizeNarrowing the sector angle may increase the frame rate. This action will increase the TI. Pulse voltag
Chapter 1 Safety 1-25 Color Sector DepthDeeper color sector depth may automatically decrease color frame rate or select a new color focal zone or c
User Manual1-26 Doppler, M-mode, and Color Imaging ControlsWhen a new imaging mode is selected, both the TI and the MI will change to default settin
Chapter 1 Safety 1-27WFUMB. Symposium on Safety of Ultrasound in Medicine: Conclusions and Recommendations on Thermal and Non-Thermal Mechanisms for
User Manual1-28l = skin line to measurement depth (cm)f = Center frequency of the transducer/system/mode combination(MHz)Since the ultrasonic path dur
Chapter 1 Safety 1-29PD The pulse duration (microseconds) associated with the transmit pattern giving rise to the reported value of the respective
User Manual1-30We also refer to a September 1993 analysis done by a working group of the IEC technical committee 87 and prepared by K. Beissner, as a
Chapter 1 Safety 1-31 Environmental ProtectionCAUTION: XThe console and peripherals could be sent back to manufacturers for recycling or proper di
SAMSUNG MEDISON DIAGNOSTIC ULTRASOUND SYSTEMVersion 2.01EnglishSONOACE R3User Manual
Chapter 2Specifications ...2-3Product Configuration ...2-5Monitor
Chapter 2 Introduction 2-3 SpecicationsPhysical DimensionsHeight: 375mm (with handle)Width: 402mm (with probe holder)Depth: 188mm (with control pan
User Manual2-4ApplicationGynecology, Abdomen, OB, Renal, Urology, Vascular, Small Part, Fetal Heart, Breast, Musculoskeletal, Pediatric, Neonatal, Car
Chapter 2 Introduction 2-5 Product CongurationThe SONOACE R3 consists of the monitor, the control panel, the console, the peripheral devices, and
User Manual2-6 Title AreaThis section displays the Logo, Patient Name, Hospital Name, Application, Frame Rate & Depth, Probe Information, Acoust
Chapter 2 Introduction 2-7Tips!Principles of Operation of the Diagnostic Ultrasound SystemMedical ultrasound images are created when the computer&ap
User Manual2-8Control PanelThe control panel can be used for controlling the system. [Figure 2.2 Control Panel]The user can manipulate the control pan
Chapter 2 Introduction 2-9 Control Panel MapThe following are descriptions and instructions for the controls on the control panel. For more inform
User Manual2-10SetUse this option to select an item or value. In Spectral Doppler Mode and 3D mode, it is also used as update function. ExitExits the
Chapter 2 Introduction 2-11 KeyboardsThis product comes with an alphanumeric keyboard, which can be used to enter text and run a variety of functi
User Manual2-12UtilityThis button is used to appear the utility menu. App.This is used to appear Application screen to select/change probes and applic
Chapter 2 Introduction 2-13ConsoleWhilst the inside of the console consists of devices enabling the implementation of ultrasound images, the outside
User Manual2-14 Rear PanelA monitor, printer, VCR, etc. are connected via the rear panel at the back of the system. 1236745[Figure 2.6 Rear Panel]1U
Chapter 2 Introduction 2-15 Power Connection PartThe power connection part is located at the back of the system. [Figure 2.7 Power Connection Part
User Manual2-16Peripheral DevicesBased on needs, various peripheral devices can be connected via the corresponding ports located on the rear panel.CAU
Chapter 2 Introduction 2-17 Inkjet PrinterNOTE: Contact Samsung Medison’s customer service department about abailable printers.CAUTION: Contact Sa
User Manual2-18ProbeProbes are devices that generate ultrasound waves and process reected wave data for the purpose of image formation.NOTE: For more
Chapter 2 Introduction 2-19SONOACE R3 Cart (Optional)[Figure 2.10 SONOACE R3 Cart]The SONOACE R3 System can be placed on a cart during use or for tr
PROPRIETARY INFORMATION AND SOFTWARE LICENSEThe Customer shall keep condential all proprietary information furnished or disclosed to the Customer by
User Manual2-20Accessory This system provides a box including accessories.CABLE CORERCA-BNC GENDER FUSEBRACKET DV12RGB GENDERDRIVERM3*8 SCREWMANUALPOW
Chapter 2 Introduction 2-21OptionsThe SONOACE R3 has the following optional functions:XImage ling-SONOVIEW XDICOMXFreehand 3DXPulse wave Dopple
Chapter 3Power Supply ...3-3Powering On ...
Chapter 3 Starting Diagnosis 3-3 Power SupplyBoot the system for use.CAUTION: Make sure to connect the probe and peripheral devices that will be u
User Manual3-4 Probes & ApplicationsBefore scanning, select a probe and an application. CAUTION: Please refer to 'Chapter 9. Probes' fo
Chapter 3 Starting Diagnosis 3-5Probe and Application SelectionSet a probe and an application in the Applications screen.1. Select a probe and an a
User Manual3-6Editing BodyMarker Select some BodyMarkers to use in the system or change the order of BodyMarkers. Click Edit BodyMarker in the Applica
Chapter 3 Starting Diagnosis 3-7 Back to default BodyMarkerPress the Default on ‘User dened BodyMarker preset’ to back to default BodyMarker.
User Manual3-8Calc Sequence[Figure 3.3 Calc Package Sequence Change]To change the order and content of the Calculation Package of each application, cl
Chapter 3 Starting Diagnosis 3-9 Patient InformationPress the Patient key on the control panel and then the Patient Information screen will be dis
User Manual3-10Registering a New Patient1. After clicking the New in the Patient Information screen, enter the patient name, ID and other information
Chapter 3 Starting Diagnosis 3-11 GA It indicates the gestational age of a patient.You can enter it manually in the specied format, or have it au
User Manual3-12 HeightEnter the patient’s height in inches or centimeters. Click the unit button to toggle the unit. If the unit is changed, the numb
Chapter 3 Starting Diagnosis 3-13Searching for Patient InformationYou can search the patient information with following three methods. Finding P
User Manual3-143. After selecting ID, press Select or Ok to display information of the selected patient. To stop the search or close the current sear
Chapter 3 Starting Diagnosis 3-15[Figure 3.6 Modality Worklist]
User Manual3-16Modifying Patient InformationTo modify information in the Patient Information screen:1. Modify the information using the trackball, Set
Chapter 4Information ... 4-3Diagnosis Mode Types ...
Safety Requirements Classication: XType of protection against electrical shock: Class IXDegree of protection against electrical shock (Patient c
Chapter 4 Diagnosis Modes4-3 InformationDiagnosis Mode TypesThis product supports a variety of diagnosis modes including Basic Mode, Combined Mode, M
User Manual4-4Basic Use The items that can be used commonly in each diagnosis mode are shown below: Gain Use the Gain dial-button on the control pane
Chapter 4 Diagnosis Modes4-51. Press the Zoom key on the keyboards.2. Use the Trackball to move the Zoom box. You can locate the Zoom box in an image
User Manual4-6 Harmonic Press the HAR key on the keyboards to select On or O. ‘HGen’, ‘HRes’. ‘HPen’ mark is displayed in the image information area
Chapter 4 Diagnosis Modes4-7 Basic Modes2D ModeThis basic mode, also referred to as B Mode (Brightness mode), provides scan planes of organs. This i
User Manual4-8 2D Mode Menu Following items are displayed on menu when you press the Menu dial-button in 2D mode. Select the item and adjust the val
Chapter 4 Diagnosis Modes4-9 Frame RateThis is the rate by which the image displayed is updated per second. Use the Menu dial-button to select Fast,
User Manual4-10 2D PostBrings up the sub-menu. [Figure 4.2 2D Post]▶2D Map: Select the post map to use. Select a map between Type 1 - 5 or User 1 –
Chapter 4 Diagnosis Modes4-11Tips!Editing a post mapYou can edit the user post map with following two methods. 1. Select the each point of the map w
Commentaires sur ces manuels