Operation Manual
5-22
Common Measurement Methods
This section provides information on the common measurement methods used for applications.
Measurements in Spectral Doppler Mode
In general, if you trace a Doppler spectrum, you can obtain results for various measurement items auto-
matically. There are 3 ways to trace a Doppler spectrum.
Auto Trace
A spectrum is traced automatically. It is enabled in the Measurement menu in Spectral Doppler Mode.
1. Press Auto Trace in the Measurement menu.
2. The system traces a spectrum automatically.
3. When Trace is complete, the measurement results are displayed on the screen.
Limited Trace
If you specify a measurement range, a spectrum is traced automatically. It is enabled in the menu in
Spectral Doppler Mode.
1. Press Limited Trace in the Measurement menu. A bar appears allowing you to specify a measure-
ment area.
2. Specify the measurement range.
Place the bar at a desired position with the trackball, and press the Set button.
3. The system traces spectrums within the specied range automatically.
4. When Trace is complete, the measurement results are displayed on the screen.
Manual Trace
A spectrum is traced manually. It is enabled in the menu in Spectral Doppler Mode.
1. Press Manual Trace in the Measurement menu. A measurement cursor appears over a spectrum.
2. Trace the spectrum. The measuring method is the same as in “D Trace.
3. When Trace is complete, the measurement results are displayed on the screen.