Operation Manual
7-38
Editing Biopsy guide line
NOTE:
X
Note that the Biopsy Guide Line cannot be edited when the Trapezoidal function is in use for the
Linear Probe.
X
If the system is rebooted, the Biopsy Guide Line settings are restored to the default.
1. Select Edit in the Biopsy menu. The warning message will appear.
2. If you click Ok, the Biopsy Edit screen appears. Click Cancel to escape from the Biopsy menu.
3. After the Biopsy Edit screen appears, you can adjust the Biopsy Guide Line with the soft menu on the
keyboard.
4. [1] Start: Move the trackball up and down to designate the starting point of the Biopsy Guide Line.
5. [2] Angle: Move the trackball right and left to decide the direction of Biopsy Guide Line movement.
6. Press the Softmenu button [4] Save to setting. The conrmation message will appear before saving
settings.
7. Click OK on the message to apply the changed Biopsy Guide Line. Press the Softmenu button [5]
Cancel to cancel saving.
[Figure 7.27 Biopsy Edit]