Samsung HS50 Manuel d'utilisateur Page 163

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 282
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 162
8-19
Chapter 8 Measurements and Calculations
Limited Trace
This is a basic measurement that is only available in Spectral Doppler mode. Specify two bars
on a Spectral Doppler image to calculate the mean velocity between the bars, RI and Pulsatility
Index(PI).
1. Select Limited Trace on the monitor or touch screen. Use the trackball and the Set button on
the control panel to select two bars.
2. When the measurement has been taken, the result is shown on the screen.
X
PS: Peak Systolic Velocity
X
MD: Min Diastolic Velocity
X
ED: End Diastolic Velocity
X
TAmax: Time Averaged Max Velocity
X
TAMEAN: Time Average Mean Velocity
X
PI: Pulsatility Index
X
RI: Resistivity Index
X
PS/ED: PS ED Ratio
X
ED/PS: ED PS Ratio
X
HR
The equations used for D Limited Trace measurement are as follows:
X
Duration of flow
VTI
meanV =
X
PS
EDPS
RI
=
X
Vmean
EDPS
PI
=
X
2
1
21
/
V
V
VV =