9-19
Chapter 9 Probes
Ultrasound transmission Gel
For proper transmission of the acoustic beam, only use ultrasound transmission gel only approved by Samsung
Medison.
WARNING:
X
Using an unsuitable ultrasound gel may damage the probe.
X
Using a damaged probe may cause an electrical shock or other injury to the patient and/or
operator. Do not use ultrasound gels that contain one or more of the following substances.
X
Do not use ultrasound gels or contact media that contain one or more of the following
substances.
Mineral oil, cooking oil, petroleum, solvents, corrosion inhibitor, lanoline paraffin, grease, ester,
excessive silicone, release agent or other oils
Acetone, methanol, plasticizer (dioctylphtalate), or other alcohols or denatured alcohols
Glacial acetic acid, iodine
All lotions or gels that contain aromatic substances