Samsung SONOACE R3 Manuel d'utilisateur Page 37

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 442
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 36
Chapter 1 Safety
1-19
Applying ALARA
The system-imaging mode used depends upon the information needed. 2D-mode and M-mode imaging
provide anatomical information, while Doppler, Power, and Color imaging provide information about
blood flow. Scanned modes, like 2D-mode, Power, or Color, disperse or scatter the ultrasonic energy
over an area, while an unscanned mode, like M-mode or Doppler, concentrates ultrasonic energy.
Understanding the nature of the imaging mode being used allows the sonographer to apply the ALARA
principle with informed judgment. The probe frequency, system set-up values, scanning techniques, and
operator experience aid the sonographer in meeting the denition of the ALARA principle.
The decision as to the amount of acoustic output is, in the nal analysis, up to the system operator. This
decision must be based on the following factors: type of patient, type of exam, patient history, ease or
diculty of obtaining diagnostically useful information, and the potential localized heating of the patient
due to probe surface temperatures. Prudent use of the system occurs when patient exposure is limited
to the lowest index reading for the shortest amount of time necessary to achieve acceptable diagnostic
results.
Although a high index reading does not mean that a bioeect is actually occurring, a high index reading
should be taken seriously. Every effort should be made to reduce the possible effects of a high index
reading. Limiting exposure time is an eective way to accomplish this goal.
There are several system controls that the operator can use to adjust the image quality and limit the
acoustic intensity. These controls are related to the techniques that an operator might use to implement
ALARA. These controls can be divided into three categories: direct, indirect, and receiver control.
Direct Controls
Application selection and the output intensity control directly aect acoustic intensity. There are dierent
ranges of allowable intensity or output based on your selection. Selecting the correct range of acoustic
intensity for the application is one of the rst things required during any exam. For example, peripheral
vascular intensity levels are not recommended for fetal exams. Some systems automatically select the
proper range for a particular procedure, while others require manual selection. Ultimately, the user bears
the responsibility for proper clinical use. The Samsung Medison system provides both automatic and user-
denable settings.
Output has direct impact on acoustic intensity. Once the application has been established, the output
control can be used to increase or decrease the intensity output. The output control allows you to select
intensity levels less than the dened maximum. Prudent use dictates that you select the lowest output
intensity consistent with good image quality.
Vue de la page 36
1 2 ... 32 33 34 35 36 37 38 39 40 41 42 ... 441 442

Commentaires sur ces manuels

Pas de commentaire