Reference Manual 119
I
PA.3
@MI
max
the derated pulse average intensity at the point of maximum reported MI
(Watts per square centimeter).
The Acoustic Measurement Precision and Acoustic Measurement Uncertainty are
summarized below.
Quantity Precision Total Uncertainty
Pll.3 (derated pulse intensity integral) 3.2 % +21 % to –24 %
Wo (acoustic power) 6.2 % +/– 19 %
Pr.3 (derated rarefaction pressure) 5.4 % +/– 15 %
Fc (center frequency) < 1 % +/– 4.5 %
Explanatory Notes
(a) This index is not required to this operating mode.
(b) This probe is not intended for adult transcranial uses.
(c) This formulation for TIS is less than that for an alternate formulation in this mode.
(d) The maximum index value is less than 1.0
* If the MI comes from a scanning mode Transmit Pattern (pulse), the ‘prr’ listed is
the average per second for the 'worst case' scan line. ‘prr’ for scanning modes is the
product of the frame rate and the number of pulse per line of the Transmit Pattern .
** The max TIS unscanned value is an ‘at_surface’ value and occurs for aperture > 1.0
cm
2
, OR The max TIS unscanned value is a ‘below_surface’ value and occurs for
aperture <= 1.0 cm
2
.
+ The max TIB for this combinational mode = the at_surface TIS_scanned value. The
non–scanned_TIB value indicated is the below_surface max. A “+” is used when the
TIB value is less than the TIS_scanned value for the operating condition.