Samsung SONOACE R5 Manuel d'utilisateur Page 410

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 454
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 409
Reference Manual 102
Eective Regurgitant Orice (ERO)
Reference: Oh, J.K., Seward, J.B., Tajik, A.J. The Echo Manual. Boston:Little, Brown and
Company, 1994, pp. 108-109
Vmax is the peak velocity of the mitral regurgitant jet measured on the Doppler display.
Regurgitant Volume (Vol)
Reference: Oh, J.K., Seward, J.B., Tajik, A.J. The Echo Manual. Boston:Little, Brown and
Company, 1994, pp. 108-109
Where:
PISA-Vol. = Regurgitant volume in milliliters
PISA-ERO = Eective Regurgitant area in cm2.
VTI = Velocity integral of the Regurgitant ow measured on the Doppler display
Cardiology Doppler
Mitral Valve, Tricuspid Valve
Pressure Gradient
where, V : Velocity (m/sec)
PHT ; Pressure Half Time
Reference : Oh, J.K., Seward, J.B., Tajik, A.J. The Echo Manual. Boston: Little, Brown and
Company, 1994, p.59-60
Pressure half-time (PHT) is the time it takes for the peak pressure gradient to fall to half to half
of its peak value. DT is deceleration time in centimeters per second form the peak velocity
to the baseline.
Mitral Valve area
where ; PHT is Pressure Half Time (milliseconds)
Reference : John H. Phillipse, “ Practical Quantitative Doppler Echocardiography”, p47,
chapter6, CRC press, 1991
Vue de la page 409
1 2 ... 405 406 407 408 409 410 411 412 413 414 415 ... 453 454

Commentaires sur ces manuels

Pas de commentaire