4-4
Operation Manual
NOTE: The functionalities for each mode may be restricted by the selected probe.
Basic Use
The items that can be used commonly in each diagnosis mode are shown below:
Using Control Panel
The items that can be used in each diagnosis mode are provided as menu items. You can change the
image format or optimize an image to facilitate your diagnosis.
■
Gain
Use the dial-button on the control panel. The Gain button appears differently depending on the
diagnosis mode, but it is usually in the form of a dial-button used to select a diagnosis mode.
You can adjust the brightness of an image. If you rotate the Gain dial-button clockwise, its value
increases.
■
TGC (Time Gain Compensation)
Use the TGC slide on the control panel.
In general, ultrasound penetration gets weaker with depth. TGC can be used to compensate for
this effect.
The product provides eight TGC slides for varying depths, allowing you to adjust Gain by area.
Among the eight slides, the top slide represents the shallowest area, while the lower slides represent
the deeper ones.
Move the slide to the right (+) to increase Gain, brightening the image.
■
Focus
Use Focus dial-button on the control panel to adjust the location of focus.
NOTE: The same dial-button is used for Focus and TB. It works as Focus in scan mode and as TB
in 3D mode.