4-12
Operation Manual
Dynamic Range
Press Dynamic Range on the touch screen. Select a Dynamic Range value between 50 - 170 by
tapping a desired button or using the dial-button on the touch screen.
Dynamic Range adjusts contrast by changing the ratio of the minimum and maximum values of
input signals. When this value becomes higher, the image is displayed more smoothly.
Frame Avg.
When an image is updated, the present image and the previous image are averaged. Tap Frame
Avg on the touch screen. Select a value between 0 and 15 by tapping the button on the touch
screen or using the dial-button. When you scan the same diagnosis site repeatedly, speckles may
appear on the updated image. Frame Avg is used to minimize these speckles.
FSI (Full Spectrum Imaging)
Use FSI on the touch screen. Select an FSI value between 1 – 3 by tapping a desired button.
This function synthesizes images with data obtained with frequencies of different characteristics.
Therefore, shallow observation depths yield higher resolution and deep observation depths yield
higher penetration. Tap FSI on the touch screen. Select a value between 1 and 3 by tapping the
button on the touch screen or using the dial-button.
Reject Level
Press the Reject Level button on the touch screen. Select a value between 1 - 32 by tapping a
desired button or using the dial-button on the touch screen.
This function is used to eliminate noise or low level echo for clearer signals.
Gray Map
Tap Gray Map on the touch screen. Select a value between 1 and 13 by tapping the button on the
touch screen or using the dial-button. This changes the 2D Post Curve.
M line
NOTE: The M line appears always in the following applications: Cardiac, TCD, and Contrast.
An M line appears on the image. The M line indicates where the observed image is located in the 2D
image when M or PW mode is used together with 2D mode.
Tap M line on the touch screen and select On or Off.