5-27
Chapter 5 Measurements and Calculations
Tips!
Things to consider for Doppler Spectrum Auto Trace
The state of a Doppler spectrum may affect measurement results. Please see the following:
Causes for Trace Failure
X
If Gain is changed for a Doppler image in the Freeze state, Contour Trace and Peak Trace will not
work.
X
If there is little or no noise in an image without a spectrum, Contour Trace will not work.
X
If there is severe noise in an image, Contour Trace will not work.
X
If the Clutter filter is set too high, Auto Trace or Limited Trace may not work.
Causes for Inaccurate Peak Trace
X
If PRF (Pulse Repetition Frequency) is lower than the velocity of the observation area, aliasing
may occur. If the original signals are separated from aliasing, Trace can be done but the peak
measurement may not be accurate.
X
If the peak of a spectral waveform is not clear or occurs intermittently, Trace can be done but the
peak measurement may not be accurate.
X
If the Doppler Gain is set to high or low, it becomes difficult to distinguish spectrums. This may
result in measurement error(s).
X
If the Wall Filter is set too high, only part of the spectrum is displayed. In this case, Trace can be
done but Peak measurement may not be accurate.
X
If abnormal noise or artifact occurs, Trace can be done but Peak measurement may not be
accurate.
Misc.
X
Use of the CW Probe may result in measurement error(s).
X
Limited Trace is supported only for two-peak spectrums such as Mitral Valve Inflow and Tricuspid
Valve Inflow in the cardiology application.
Limited Trace
If you specify a measurement range, a spectrum is traced automatically. It is enabled in the menu
in Spectral Doppler Mode.
1. Press Limited Trace in the Measurement menu. A bar appears allowing you to specify a
measurement area.
2. Specify the measurement range.
X
Place the bar at a desired position with the trackball, and press the Set button.
3. The system traces spectrums within the specied range automatically.
4. When Trace is complete, the measurement results are displayed on the screen.